Detcon DM-700-O2 Oxygen Sensor

  1. Home
  2. Products

Detcon DM-700-O2 Oxygen Sensor

The Detcon DM-700-O2 by Teledyne is a non-intrusive “Smart” sensor designed to detect and monitor Oxygen in the air over the range of 0-25% by volume using electrochemical sensor technology. The plug-in, field-replaceable sensor cell is temperature compensated and features oversized gold-plated connections that help prevent corrosion.

The Model 700’s rugged framework includes an intrinsically safe electro-polished 316 stainless steel housing with fully encapsulated electronics and dual-layer surge protection. This innovative design virtually eliminates sensor failure due to water ingress, corrosion, vibration, and transient spikes.

A primary feature of the Model 700 is embedded intuitive software that simplifies the operator interface by guiding the user through routine calibration, configuration, and fault diagnostic functions using a built-in alpha/numeric display. The Model 700 is equipped with standard analogue 4-20mA and Modbus RS-485 outputs. Among its unique features is a wireless option that can be used with Detcon’s Smart Wireless product line.

Additional integration options include a Remote Alarm Module (RAM) and HART. Detcon’s oxygen sensors have a long shelf life and are supported by an industry-leading warranty.

Applications: Petrochemical, Refineries, Chemical plants, Confined Space Monitoring, Analyser Shelters, Wastewater treatment plants, Wet wells-dry wells, Pulp and paper, Food and beverage and Utilities.

Features

Failsafe User-Friendly Interface

  • LED Display (With Antiglare Cover)
  • Full-Text Display Method
  • Non-intrusive Interface
  • Auto Zero/Auto Span
  • Pre-emptive Fault Diagnostics

Environmentally Bulletproof

  • Electropolished 316SS Construction
  • 100% Epoxy Encapsulated Circuitry
  • Bulletproof I/O Protection
  • Water-Proof, Corrosion-Proof
  • Vibration-Proof

Modular and Serviceable

  • Modular Design
  • Plug and Play Components
  • Quick Thread Release (For Sensor Replacement)
  • Integral Calibration Port

Order Information

700 Series Oxygen Sensor Assemblies

Part Number Description
967-343490-25P DM-700-O2 Oxygen Deficiency Sensor Assembly (no condulet)
967-343491-25P DM-700-O2 Oxygen Deficiency Gas Sensor Assembly (aluminium condulet)
967-343494-25P DM-700-O2 Oxygen Deficiency Gas Sensor Assembly (316SS condulet)
9H7-343491-25P DM-700-O2 Oxygen Deficiency Gas Sensor with HART bridge Assembly (aluminium condulet)
9H7-343494-25P DM-700-O2 Oxygen Deficiency Gas Sensor with HART bridge Assembly (316SS condulet)
9R7-343491-25P DM-700-O2 Oxygen Deficiency Gas Sensor with RAM Assembly (aluminium condulet)
9R7-343494-25P DM-700-O2 Oxygen Deficiency Gas Sensor with RAM Assembly (316SS condulet)
9K7-343491-25P DM-700-O2 Oxygen Deficiency Gas Sensor with HART-RAM Assembly (aluminium condulet)
9K7-343494-25P DM-700-O2 Oxygen Deficiency Gas Sensor with HART-RAM Assembly (316SS condulet)

Spares

Part Number Description
377-343401-025 O2 Oxygen plug-in sensor for DM-700
S927-343400-25P DM-700-O2 Oxygen Deficiency (ITM)
602-003803-000 CXT / 700 DM Splashguard Adapter (Stainless Steel)
System Specifications
Sensor Type Continuous diffusion/adsorption, 2-electrode electrochemical cell, plug-in field replaceable intelligent type
Sensor Life 2.5 to 3 years typical
Measurement Range 0-25% Oxygen by Volume
Accuracy / Repeatability ±1% FS
Response / Clearing Time T90 ≤ 10 seconds
Span Drift <5% of signal per year (in first two years)
Outputs Linear 4-20 mA DC, RS-485 Modbus RTU
Electrical Classification Explosion Proof
CSA and US (NRTL)
Class I, Division 1, Groups B, C, D (Tamb = -40°C to +50°C)
Class I, Zone 1, Group IIC
ATEX: II 2 G Ex d ib IIB + H2 T4 Gb (Tamb = -40°C to +50°C)
Ingress Protection NEMA 4X, IP66
Safety Approvals cCSAus, ATEX, CE Marking, SIL2 Certified to IEC 61508
Warranty Plug-in Cell – 2 Years; Transmitter – 2 Years
Environmental Specifications
Operating Temperature Range -40°F to +122°F (-40°C to +50°C)
Storage Temperature Range -40°F to +122°F (-40°C to +50°C)
Operating Humidity Range 0-99% RH Non-Condensing
Operating Pressure Range Atmospheric ±10%
Mechanical Specifications
Dimensions Sensor Assembly: 7" H × 2.2" Dia. (178 mm H × 65 mm Dia.)
With Junction Box: 11" H × 6.1" W × 3.75" D (280 mm H × 155 mm W × 96 mm D)
Mounting Holes (J-Box): 5.5" (140 mm) center-to-center
Weight 2 lbs (0.907 kg) – Sensor Only
6 lbs (2.72 kg) – With Aluminum Junction Box
9 lbs (4.08 kg) – With Stainless Steel Junction Box
Electrical Specifications
Power Input 11-30 VDC
Power Consumption Normal Operation: 30 mA (<0.75 W)
Maximum: 50 mA (1.2 W)
Inrush Current 1.67 A @ 24 VDC
RFI / EMI Protection Complies with EN50270
Analog Output Linear 4-20 mA DC (1,000 Ω max loop load @ 24 VDC)
0 mA – All Fault Diagnostics
2 mA – In Calibration
4-20 mA – 0-100% Full Scale
22 mA – Over-Range Condition
Serial RS-485 Output RS-485 Modbus™ RTU
Baud Rate 9600 BPS (9600, N, 8, 1 Half Duplex)
Status Indicators 4-digit LED Display with Gas Concentration, Full-Script Menu Prompts for AutoSpan, Set-up Options and Fault Reporting
Faults Monitored Loop, Input Voltage, Missing Sensor, Zero, Processor, Memory, Calibration
Cable Requirements Power/Analog: 3-wire shielded cable, maximum distance 13,300 ft with 14 AWG
Serial Output: 2-wire twisted-pair shielded cable for RS-485 installations, maximum distance 4,000 ft to last sensor
I/O Protection Over-voltage, Miswiring, EMI/RFI Immunity